In the framework of the realization of dedicated Cone-Beam Breast Computed Tomography (CBBCT) apparatus we report the MTF characterization of our prototype having the best performance in terms of detector pixel size and X ray focal spot than any other systems built or under construction. From our measurement we estimated a value of the longitudinal and transverse resolution of 4-5 lp/mm (at 10% MTF). The presence of CT image features has been observed and attributed of specific phase contrast effects.
Measurement of the MTF of a Cone-Beam Breast Computed Tomography laboratory scanner / Russo, Paolo; Mettivier, Giovanni. - (2009). (Intervento presentato al convegno IEEE Medical Imaging Conference tenutosi a Orlando, FLorida, USA nel 25-31 October 2009).
Measurement of the MTF of a Cone-Beam Breast Computed Tomography laboratory scanner.
RUSSO, PAOLO;METTIVIER, GIOVANNI
2009
Abstract
In the framework of the realization of dedicated Cone-Beam Breast Computed Tomography (CBBCT) apparatus we report the MTF characterization of our prototype having the best performance in terms of detector pixel size and X ray focal spot than any other systems built or under construction. From our measurement we estimated a value of the longitudinal and transverse resolution of 4-5 lp/mm (at 10% MTF). The presence of CT image features has been observed and attributed of specific phase contrast effects.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.