We present in vivo tomographic 125I imaging of a mouse thyroid carried with MediSPECT, a scanner developed at University & INFN Napoli, for SPECT studies on small animals with a small field of view (FoV) and high spatial resolution. MediSPECT is based on a fine pitch CdTe pixel detector, operating in single photon counting, (14×14 mm2, 256×256 square pixel of 55 micrometer side) and equipped with a high resolution collimators set. The scanner is composed by a gantry, hosting the collimation and detection units, rotating around a horizontal axis along which is positioned the small animal holder. The mouse has been injected with a Na125I solution with a total activity of 31.8 MBq. The planar projections for SPECT reconstruction have been acquired using a 300 m tungsten pinhole (magnification 1.47 and a FoV of 9.6×9.6 mm2). The projections have been cap-tured in a step-and shoot fashion and their SPECT reconstruc-tion has been done with a C software based on Ordered Subsets-Expectation Maximization (OS-EM) reconstruction algorithm. The spatial resolution with a 300 micrometer inhole is, in planar imaging, about 500 micrometer and, in tomographic imaging, better than 1 mm. The MediSPECT scanner has a modular structure that allows for multimodal imaging: at present, a Fluorescence Reflectance Imaging head has been installed, for planar optical imaging in addition to SPECT imaging.

The MediSPECT Small Animal CdTe Scanner / Russo, Paolo. - (2009). (Intervento presentato al convegno World Congress on Medical Physics and Biomedical Engineering tenutosi a Munich, Germany nel 7-12 settembre 2009).

The MediSPECT Small Animal CdTe Scanner.

RUSSO, PAOLO
2009

Abstract

We present in vivo tomographic 125I imaging of a mouse thyroid carried with MediSPECT, a scanner developed at University & INFN Napoli, for SPECT studies on small animals with a small field of view (FoV) and high spatial resolution. MediSPECT is based on a fine pitch CdTe pixel detector, operating in single photon counting, (14×14 mm2, 256×256 square pixel of 55 micrometer side) and equipped with a high resolution collimators set. The scanner is composed by a gantry, hosting the collimation and detection units, rotating around a horizontal axis along which is positioned the small animal holder. The mouse has been injected with a Na125I solution with a total activity of 31.8 MBq. The planar projections for SPECT reconstruction have been acquired using a 300 m tungsten pinhole (magnification 1.47 and a FoV of 9.6×9.6 mm2). The projections have been cap-tured in a step-and shoot fashion and their SPECT reconstruc-tion has been done with a C software based on Ordered Subsets-Expectation Maximization (OS-EM) reconstruction algorithm. The spatial resolution with a 300 micrometer inhole is, in planar imaging, about 500 micrometer and, in tomographic imaging, better than 1 mm. The MediSPECT scanner has a modular structure that allows for multimodal imaging: at present, a Fluorescence Reflectance Imaging head has been installed, for planar optical imaging in addition to SPECT imaging.
2009
The MediSPECT Small Animal CdTe Scanner / Russo, Paolo. - (2009). (Intervento presentato al convegno World Congress on Medical Physics and Biomedical Engineering tenutosi a Munich, Germany nel 7-12 settembre 2009).
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11588/412531
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 0
  • ???jsp.display-item.citation.isi??? 0
social impact