We present the first tests of a digital imaging system based on a silicon pixel detector bump bonded to an integrated circuit operating in single-photon counting mode. The X-ray sensor is a 300-um-thick silicon, 14x14 mm2 sensitive area, upon which a matrix of 256x256 (65536) square pixel, 55 um in side, has been built. The readout chip, named MEDIPIX2, has been developed at CERN within the MEDIPIX2 collaboration and it is composed by a matrix of 65536 readout channels exactly with the same geometry of the detector. The spatial resolution properties of the system have been assessed by measuring the square-wave resolution function and the first images of a standard mammographic phantom were acquired using a radiographic tube in the clinical irradiation condition.

A Medipix2-based imaging system for digital mammography with silicon pixel detectors / Bisogni, M. G.; Delogu, P.; Fantacci, M. E.; Mettivier, Giovanni; Montesi, MARIA CRISTINA; Novelli, M.; Quattrocchi, M.; Rosso, V.; Russo, Paolo; Stefanini, A.. - In: IEEE TRANSACTIONS ON NUCLEAR SCIENCE. - ISSN 0018-9499. - STAMPA. - 51:(2004), pp. 3081-3085. [10.1109/TNS.2004.839079]

A Medipix2-based imaging system for digital mammography with silicon pixel detectors

METTIVIER, GIOVANNI;MONTESI, MARIA CRISTINA;RUSSO, PAOLO;
2004

Abstract

We present the first tests of a digital imaging system based on a silicon pixel detector bump bonded to an integrated circuit operating in single-photon counting mode. The X-ray sensor is a 300-um-thick silicon, 14x14 mm2 sensitive area, upon which a matrix of 256x256 (65536) square pixel, 55 um in side, has been built. The readout chip, named MEDIPIX2, has been developed at CERN within the MEDIPIX2 collaboration and it is composed by a matrix of 65536 readout channels exactly with the same geometry of the detector. The spatial resolution properties of the system have been assessed by measuring the square-wave resolution function and the first images of a standard mammographic phantom were acquired using a radiographic tube in the clinical irradiation condition.
2004
A Medipix2-based imaging system for digital mammography with silicon pixel detectors / Bisogni, M. G.; Delogu, P.; Fantacci, M. E.; Mettivier, Giovanni; Montesi, MARIA CRISTINA; Novelli, M.; Quattrocchi, M.; Rosso, V.; Russo, Paolo; Stefanini, A.. - In: IEEE TRANSACTIONS ON NUCLEAR SCIENCE. - ISSN 0018-9499. - STAMPA. - 51:(2004), pp. 3081-3085. [10.1109/TNS.2004.839079]
File in questo prodotto:
File Dimensione Formato  
a medipix2-based.pdf

non disponibili

Tipologia: Documento in Post-print
Licenza: Accesso privato/ristretto
Dimensione 668.64 kB
Formato Adobe PDF
668.64 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11588/110390
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 19
  • ???jsp.display-item.citation.isi??? 18
social impact