The paper presents a digital apparatus for measuring current transformer errors in the range of frequencies between 50 and 1,250 Hz employing wattmetrical techniques: ratio and angle errors are calculated with accuracies of 10** minus **4 and 0. 1 prime respectively. Experimental results show the influence of different parameters which affect the measurement's accuracy; it the possibility of improving the accuracy of the device in the same range of frequency is foreseen.
Digital Equipment for Testing Current Transformers in the Range of Audio Frequencies. [APPARECCHIATURA NUMERICA PER LA VERIFICA DEGLI ERRORI DEI TRASFORMATORI DI CORRENTE NEL CAMPO DELLE FREQUENZE ACUSTICHE.] / D'Apuzzo, Massimo; M., Savastano. - In: L'ENERGIA ELETTRICA. - ISSN 1590-7651. - STAMPA. - 59:(1982), pp. 15-21.
Digital Equipment for Testing Current Transformers in the Range of Audio Frequencies. [APPARECCHIATURA NUMERICA PER LA VERIFICA DEGLI ERRORI DEI TRASFORMATORI DI CORRENTE NEL CAMPO DELLE FREQUENZE ACUSTICHE.]
D'APUZZO, MASSIMO;
1982
Abstract
The paper presents a digital apparatus for measuring current transformer errors in the range of frequencies between 50 and 1,250 Hz employing wattmetrical techniques: ratio and angle errors are calculated with accuracies of 10** minus **4 and 0. 1 prime respectively. Experimental results show the influence of different parameters which affect the measurement's accuracy; it the possibility of improving the accuracy of the device in the same range of frequency is foreseen.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.