BACKGROUND: Dental procedures involve contact between instruments and the patient's tissues, blood or saliva. This study evaluated the efficacy of the standardized sterilization of non-disposable air/water syringe tips and corrosion and contaminant build-up in these tips. METHODS: The bacterial contamination of single-use and multiple-use non-disposable air/water syringe tips after routine use and sterilization was compared to that of single-use disposable tips by microbial culturing on PCA and blood agar plates. The effect of flushing the syringe tips prior to sterilization was also measured. The amount of corrosion in single-use and multiple-use non-disposable syringes was measured by SEM and EDS analyses. RESULTS: Non-disposable syringe tips had significantly (p < 0.05) greater bacterial contamination than single-use disposable tips. There were no statistically different levels of contamination between flushed and non-flushed non-disposable syringes or between single-use and multiple-use non-disposable syringes. SEM and EDS analyses showed greater evidence of corrosion and contaminant build-up in multiple-use syringes compared to single-use non-disposable syringes. CONCLUSIONS: Sterilization of non-disposable air/water syringes is not completely effective and rinsing, or the number of uses, does not affect the effectiveness of sterilization. There may be a lower risk of cross-infection from the use of disposable air/water syringe tips, instead of non-disposable ones.
Efficacy of air/water syringe tip sterilization / Inger, M; Bennani, V; Farella, Mauro; Bennani, F; Cannon, R.. - In: AUSTRALIAN DENTAL JOURNAL. - ISSN 0045-0421. - (2014). [10.1111/adj.12146]
Efficacy of air/water syringe tip sterilization.
FARELLA, MAURO;
2014
Abstract
BACKGROUND: Dental procedures involve contact between instruments and the patient's tissues, blood or saliva. This study evaluated the efficacy of the standardized sterilization of non-disposable air/water syringe tips and corrosion and contaminant build-up in these tips. METHODS: The bacterial contamination of single-use and multiple-use non-disposable air/water syringe tips after routine use and sterilization was compared to that of single-use disposable tips by microbial culturing on PCA and blood agar plates. The effect of flushing the syringe tips prior to sterilization was also measured. The amount of corrosion in single-use and multiple-use non-disposable syringes was measured by SEM and EDS analyses. RESULTS: Non-disposable syringe tips had significantly (p < 0.05) greater bacterial contamination than single-use disposable tips. There were no statistically different levels of contamination between flushed and non-flushed non-disposable syringes or between single-use and multiple-use non-disposable syringes. SEM and EDS analyses showed greater evidence of corrosion and contaminant build-up in multiple-use syringes compared to single-use non-disposable syringes. CONCLUSIONS: Sterilization of non-disposable air/water syringes is not completely effective and rinsing, or the number of uses, does not affect the effectiveness of sterilization. There may be a lower risk of cross-infection from the use of disposable air/water syringe tips, instead of non-disposable ones.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.