Purpose: To evaluate the impact on dose distribution to eye organs-at-risk (eOARs) of a computed tomography (CT)-based treatment planning in eye plaque brachytherapy (EPB) treatment. Methods: We analyzed 19 ocular melanoma patients treated with ruthenium-106 plaques to a total dose of 100 Gy to tumor apex using conventional central-axis-point dose calculation. Treatments were re-planned using the Plaque Simulator (PS) software implementing two different strategies: a personalized CT-eye-model (CT-PS) and a standard-eye-model (SEM-PS) defined by Collaborative Ocular Melanoma Study. Dice coefficient and Hausdorff distance evaluated the concordance between eye-bulb-models. Mean doses (Dmean) to tumor and eOARs were extracted from Dose-Volume-Histograms and Retinal-Dose-Area-Histogram. Differences between planning approaches were tested by Wilcoxon signed-rank test. Results: In the analyzed cohort, 8 patients (42%) had posterior tumor location, 8 (42%) anterior, and 3 (16%) equatorial. The SEM did not accurately described the real CT eye-bulb geometry (median Hausdorff distance 0.8 mm, range: (0.4–1.3) mm). Significant differences in fovea and macula Dmean values were found (p = 0.04) between CT-PS and SEM-PS schemes. No significant dosimetric differences were found for tumor and other eOARs. The planning scheme particularly affects the OARs closest to the tumor with a general tendency of SEM-PS to overestimate the doses to the OARs closest to the tumor. Conclusion: The dosimetric accuracy achievable with CT-PS EPB treatment planning may help to identify ocular melanoma patients who could benefit the most from a personalized eye dosimetry for an optimal outcome in terms of tumor coverage and eOARs sparing. Further research and larger studies are underway.
Personalized treatment planning in eye brachytherapy for ocular melanoma: Dosimetric analysis on ophthalmic structure at risk / Itta, F.; Liuzzi, R.; Farella, A.; Porri, G.; Pacelli, R.; Conson, M.; Oliviero, C.; Buonanno, F.; Breve, M. A.; Cennamo, G.; SERVODIO IAMMARRONE, Clemente; Cella, L.. - In: PHYSICA MEDICA. - ISSN 1120-1797. - 76:(2020), pp. 285-293. [10.1016/j.ejmp.2020.07.019]
Personalized treatment planning in eye brachytherapy for ocular melanoma: Dosimetric analysis on ophthalmic structure at risk
Farella A.;Porri G.;Pacelli R.;Conson M.;Oliviero C.;Buonanno F.;Breve M. A.;Cennamo G.;Clemente S.;Cella L.
2020
Abstract
Purpose: To evaluate the impact on dose distribution to eye organs-at-risk (eOARs) of a computed tomography (CT)-based treatment planning in eye plaque brachytherapy (EPB) treatment. Methods: We analyzed 19 ocular melanoma patients treated with ruthenium-106 plaques to a total dose of 100 Gy to tumor apex using conventional central-axis-point dose calculation. Treatments were re-planned using the Plaque Simulator (PS) software implementing two different strategies: a personalized CT-eye-model (CT-PS) and a standard-eye-model (SEM-PS) defined by Collaborative Ocular Melanoma Study. Dice coefficient and Hausdorff distance evaluated the concordance between eye-bulb-models. Mean doses (Dmean) to tumor and eOARs were extracted from Dose-Volume-Histograms and Retinal-Dose-Area-Histogram. Differences between planning approaches were tested by Wilcoxon signed-rank test. Results: In the analyzed cohort, 8 patients (42%) had posterior tumor location, 8 (42%) anterior, and 3 (16%) equatorial. The SEM did not accurately described the real CT eye-bulb geometry (median Hausdorff distance 0.8 mm, range: (0.4–1.3) mm). Significant differences in fovea and macula Dmean values were found (p = 0.04) between CT-PS and SEM-PS schemes. No significant dosimetric differences were found for tumor and other eOARs. The planning scheme particularly affects the OARs closest to the tumor with a general tendency of SEM-PS to overestimate the doses to the OARs closest to the tumor. Conclusion: The dosimetric accuracy achievable with CT-PS EPB treatment planning may help to identify ocular melanoma patients who could benefit the most from a personalized eye dosimetry for an optimal outcome in terms of tumor coverage and eOARs sparing. Further research and larger studies are underway.File | Dimensione | Formato | |
---|---|---|---|
Physica Medica.pdf
accesso aperto
Tipologia:
Documento in Post-print
Licenza:
Dominio pubblico
Dimensione
1.8 MB
Formato
Adobe PDF
|
1.8 MB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.