Purpose: To report the level of agreement, repeatability, and correlation of axial length (AL), mean keratometry, central corneal thickness, anterior chamber depth, lens thickness, and corneal diameter measurements of 2 swept-source optical coherence tomography (SS-OCT) biometers, the IOLMaster 700 (reference biometer) and the Argos (new biometer). Setting: Moorfields Eye Hospital NHS Foundation Trust, London, United Kingdom. Design: Retrospective case series. Methods: Each patient had SS-OCT biometry with the reference biometer and new biometer. In addition to reporting the statistical results derived from both eyes, this study included a subgroup analysis of right eyes and left eyes. The level of agreement between the biometers was represented with the Bland-Altman method. Power vector analysis of the J0 (Jackson cross-cylinder, axes at 0 degrees and 90 degrees) and J45 (Jackson cross-cylinder, axes at 45 degrees and 135 degrees) vectorial components of astigmatism was performed. Internal consistency was assessed with the Cronbach α coefficient of reliability. The dispersion of probability distribution was computed with the coefficient of variation. The intraoperator repeatability was calculated using the intraclass correlation coefficient (ICC). Results: The study comprised 112 patients (218 eyes). There was a statistically significant difference between the 2 biometers in all measurements (P < .05) except AL. The level of correlation between the reference biometer and the new biometer was very high for all the parameters except corneal diameter, and the agreement was high. The ICC and internal consistency were excellent with both biometers. Conclusion: The new biometer provided good agreement and repeatability compared with the reference biometer.

Comparative analysis of 2 swept-source optical coherence tomography biometers / Sabatino, F.; Matarazzo, F.; Findl, O.; Maurino, V.. - In: JOURNAL OF CATARACT AND REFRACTIVE SURGERY. - ISSN 0886-3350. - 45:8(2019), pp. 1124-1129. [10.1016/j.jcrs.2019.03.020]

Comparative analysis of 2 swept-source optical coherence tomography biometers

Matarazzo F.
Secondo
;
2019

Abstract

Purpose: To report the level of agreement, repeatability, and correlation of axial length (AL), mean keratometry, central corneal thickness, anterior chamber depth, lens thickness, and corneal diameter measurements of 2 swept-source optical coherence tomography (SS-OCT) biometers, the IOLMaster 700 (reference biometer) and the Argos (new biometer). Setting: Moorfields Eye Hospital NHS Foundation Trust, London, United Kingdom. Design: Retrospective case series. Methods: Each patient had SS-OCT biometry with the reference biometer and new biometer. In addition to reporting the statistical results derived from both eyes, this study included a subgroup analysis of right eyes and left eyes. The level of agreement between the biometers was represented with the Bland-Altman method. Power vector analysis of the J0 (Jackson cross-cylinder, axes at 0 degrees and 90 degrees) and J45 (Jackson cross-cylinder, axes at 45 degrees and 135 degrees) vectorial components of astigmatism was performed. Internal consistency was assessed with the Cronbach α coefficient of reliability. The dispersion of probability distribution was computed with the coefficient of variation. The intraoperator repeatability was calculated using the intraclass correlation coefficient (ICC). Results: The study comprised 112 patients (218 eyes). There was a statistically significant difference between the 2 biometers in all measurements (P < .05) except AL. The level of correlation between the reference biometer and the new biometer was very high for all the parameters except corneal diameter, and the agreement was high. The ICC and internal consistency were excellent with both biometers. Conclusion: The new biometer provided good agreement and repeatability compared with the reference biometer.
2019
Comparative analysis of 2 swept-source optical coherence tomography biometers / Sabatino, F.; Matarazzo, F.; Findl, O.; Maurino, V.. - In: JOURNAL OF CATARACT AND REFRACTIVE SURGERY. - ISSN 0886-3350. - 45:8(2019), pp. 1124-1129. [10.1016/j.jcrs.2019.03.020]
File in questo prodotto:
File Dimensione Formato  
sabatino matarazzo 2019.pdf

solo utenti autorizzati

Tipologia: Versione Editoriale (PDF)
Licenza: Copyright dell'editore
Dimensione 1.06 MB
Formato Adobe PDF
1.06 MB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11588/959370
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 34
  • ???jsp.display-item.citation.isi??? 33
social impact