Electromagnetic field simulations are increasingly used to assure RF safety of patients during MRI exams. In practice, however, tissue property distribution of the patient being imaged is not known, but may be represented with a pre-existing model. Repeatedly, agreement in transmit magnetic ( B1+) field distributions between two geometries has been used to suggest agreement in heating distributions. Here we examine relative effects of anatomical differences on B1+ distribution, specific absorption rate (SAR), and temperature change (ΔT). Numerical simulations were performed for a single surface coil positioned adjacent a homogeneous phantom and bovine phantom, each with slight geometric variations, and adjacent two different human body models. Experimental demonstration was performed on a bovine phantom using MR thermometry and B1+ mapping. Simulations and experiments demonstrate that B1+ distributions in different samples can be well correlated, while notable difference in maximum SAR and ΔT occur. This work illustrates challenges associated with utilizing simulations or experiments for RF safety assurance purposes. Reliance on B1+ distributions alone for validation of simulations and/or experiments with a sample or subject for assurance of safety in another should be performed with caution.

Effects of anatomical differences on electromagnetic fields, SAR, and temperature change / Alon, L.; Deniz, C. M.; Carluccio, G.; Brown, R.; Sodickson, D. K.; Collins, C. M.. - In: CONCEPTS IN MAGNETIC RESONANCE. PART B, MAGNETIC RESONANCE ENGINEERING. - ISSN 1552-5031. - 46:1(2016), pp. 8-18. [10.1002/cmr.b.21317]

Effects of anatomical differences on electromagnetic fields, SAR, and temperature change

Carluccio G.;
2016

Abstract

Electromagnetic field simulations are increasingly used to assure RF safety of patients during MRI exams. In practice, however, tissue property distribution of the patient being imaged is not known, but may be represented with a pre-existing model. Repeatedly, agreement in transmit magnetic ( B1+) field distributions between two geometries has been used to suggest agreement in heating distributions. Here we examine relative effects of anatomical differences on B1+ distribution, specific absorption rate (SAR), and temperature change (ΔT). Numerical simulations were performed for a single surface coil positioned adjacent a homogeneous phantom and bovine phantom, each with slight geometric variations, and adjacent two different human body models. Experimental demonstration was performed on a bovine phantom using MR thermometry and B1+ mapping. Simulations and experiments demonstrate that B1+ distributions in different samples can be well correlated, while notable difference in maximum SAR and ΔT occur. This work illustrates challenges associated with utilizing simulations or experiments for RF safety assurance purposes. Reliance on B1+ distributions alone for validation of simulations and/or experiments with a sample or subject for assurance of safety in another should be performed with caution.
2016
Effects of anatomical differences on electromagnetic fields, SAR, and temperature change / Alon, L.; Deniz, C. M.; Carluccio, G.; Brown, R.; Sodickson, D. K.; Collins, C. M.. - In: CONCEPTS IN MAGNETIC RESONANCE. PART B, MAGNETIC RESONANCE ENGINEERING. - ISSN 1552-5031. - 46:1(2016), pp. 8-18. [10.1002/cmr.b.21317]
File in questo prodotto:
File Dimensione Formato  
Concepts Magn Reson - 2015 - Alon - Effects of anatomical differences on electromagnetic fields SAR and temperature.pdf

accesso aperto

Licenza: Creative commons
Dimensione 637.13 kB
Formato Adobe PDF
637.13 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11588/985328
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 28
  • ???jsp.display-item.citation.isi??? 25
social impact